Product Details:
Payment & Shipping Terms:
|
Product Name: | Double Door Autoclave | Sterilization Medium: | Moist And Heat With Pure Steam |
---|---|---|---|
Programs: | Gravity And Vacuum | Material: | 304 Chamber And Carbon Steel Jacket |
Chamber Minor Treatment: | <8 Ra | Door Type: | Sliding Single Door |
Installation: | Floor Standing Or Embedded | Capacity: | 1200 Liters |
Door Safety: | Can’t Open Under Pressure | Control: | Microprocessor |
High Light: | medical waste sterilizer,medical waste treatment equipment |
Bulk Double Door Autoclave With 5.7 Upto 8 Inches Wide Colorful LCD Display
Specifications
Model |
Chamber Size |
Overall Size |
Net Weight |
Steam |
Water |
Power |
W*H*D mm |
W*H*D mm |
Kg |
Kg/cycle |
Kg/cycle |
KVA |
|
MD-0.8Y |
680*1180*1000 |
2200*2000*1410 |
1480 |
35 |
80 |
3 |
MD-1.0Y |
680*1180*1250 |
2200*2000*1660 |
1630 |
40 |
100 |
4 |
MD-1.2Y |
680*1180*1500 |
2200*2000*1910 |
2030 |
45 |
120 |
4 |
MD-1.5Y |
680*1180*1870 |
2200*2000*2280 |
2230 |
50 |
150 |
4.5 |
MD-2.0Y |
1000*1200*1700 |
2800*2000*2110 |
2490 |
60 |
200 |
4.5 |
MD-3.0Y |
1000*1500*2000 |
2800*2200*2410 |
3050 |
90 |
300 |
5 |
MD-4.0Y |
1000*1500*2670 |
2800*2200*3080 |
3700 |
120 |
400 |
6 |
MD-5.0Y |
1000*1500*3340 |
2800*2200*3750 |
4600 |
150 |
500 |
8 |
MD-6.0Y |
1000*1500*4000 |
2800*2200*4410 |
5400 |
180 |
600 |
8 |
MD-8.0Y |
1000*1500*5340 |
2800*2200*5750 |
7000 |
240 |
800 |
12 |
Description
Applications
Double Door Autoclave are designed to focus on research center, animal laboratory and medical center for their large volume & easy operating.
Competitive Advantage
Double Door Autoclave of this model now adopted is the control of sterilization phase
according to a pressure signal, rather than temperature. Obviously, the temperature readout
varies as a function of the position of the sensors, whilst the pressure is uniform throughout
the chamber.
However, process control is performed by reading in the chamber (and from inside
the product) temperature signals whose match with the pressure value, as constantly
verified by the controller, demonstrates that the autoclave is operating correctly.
This principle, in saturated-steam conditions(when there is a two-way T/P correspondence)
allows to obtain uniformities in sterilization temperature which are far better then the currently
required level of ±1℃
Contact Person: Ms. JULIET
Tel: +8613915466171